CADTH Health Technology Review

Reprocessed Single-Use Semicritical and Critical Medical Devices

Rapid Review

Authors: Kellee Kaulback, Jennifer Horton

Key Messages

What Is the Issue?

What Did We Do?

What Did We Find?

What Does it Mean?

Abbreviations

MI

myocardial infarction

PSI

port-site infection

SUMDs

single-use medical devices

Context and Policy Issues

What Is Medical Device Reprocessing?

Reprocessing a medical device encompasses cleaning, reconditioning, function testing, and disinfection or sterilization to ensure the device can safely be reused.1 The terms reuse, reprocessing, and remanufacturing are used somewhat interchangeably in the literature. However, the terms reuse and reprocessing can often refer to more rudimentary checking, cleaning and re-sterilization, often within the health care institution.2 On the other hand, remanufacturing usually describes a process conducted by a third-party, and modern remanufacturing is performed on an industrial scale with rigorous adherence to defined protocols, producing a remanufactured device which has been performance-tested, shown to meet the original equipment manufacturer requirements, and demonstrates verifiable elimination of infective drugs.2

It is up to each manufacturer to decide how to design and label each medical device, and SUMDs that Health Canada licences are intended by their manufacturers to be used once during a single procedure and not to be disassembled, cleaned, reassembled, and reused, where doing so can jeopardize their performance, safety, and effectiveness.3 Manufacturers of SUMDs are not obliged to provide instructions for proper cleaning and sterilization, as they are for reusable devices.

Provincial and territorial policies regarding the reprocessing of medical devices are typically based on the device category, and the internationally accepted classification scheme described by Spaulding, which groups devices according to the risk of infection associated with the device, categorizes medical devices, as follows:4,5

Why Is it Important to Do This Review?

In 2016, Health Canada updated its policy on the purchase and use of reprocessed SUMDs, concluding that it has authority to require that commercially reprocessed devices meet appropriate standards for safety, effectiveness, and labelling, with third-party reprocessing companies held to the same requirements as manufacturers of new devices.6 However, Health Canada does not provide oversight for hospital onsite reprocessing or in community health care settings, deferring to the oversight provided at the provincial and territorial level. Provincial policies generally refer to using third-party commercial processors in accordance with Health Canada requirements.7,8 Given the potential economic and environmental benefits of using reprocessed SUMDs,9-12 yet limited guidance for hospitals to pursue onsite SUMD reprocessing, there is a growing interest in determining the clinical safety of reprocessed SUMDs.

In June 2023, CADTH published a brief summary of studies evaluating critical device category SUMDs that were published over the previous 5 years.5 The report provided limited abstract-level information from 3 studies that met the inclusion criteria. Therefore, this Rapid Review is being undertaken to expand upon the previous report by providing a full-text review and critical appraisal of relevant studies, without limiting to literature published within the past 5 years and including any literature in semicritical device categories. The 3 studies identified in the original report will also be critically appraised and summarized in this update.

Objective

To support decision-making about the appropriate use of reprocessed critical and semicritical SUMDs, this report aims to identify literature evaluating the clinical safety of reprocessed SUMDs compared with nonreprocessed (new) SUMDs.

Research Question

What is the clinical evidence regarding the safety of reprocessed single-use semicritical or critical medical devices compared with new single-use semicritical or critical medical devices?

Methods

Literature Search Methods

An information specialist conducted a literature search on key resources including MEDLINE, the Cochrane Database of Systematic Reviews, the International HTA Database, the websites of Canadian and major international health technology agencies, as well as a focused internet search. The search approach was customized to retrieve a limited set of results, balancing comprehensiveness with relevancy. The search strategy comprised both controlled vocabulary, such as the National Library of Medicine’s MeSH (Medical Subject Headings), and keywords. Search concepts were developed based on the elements of the research questions and selection criteria. The main search concepts were single-use medical devices and reprocessing. CADTH-developed search filters were applied to limit retrieval to health technology assessments, systematic reviews, meta-analyses, indirect treatment comparisons, and any types of clinical trials or observational studies. The search was completed on September 7, 2023.

Selection Criteria and Methods

One reviewer screened citations and selected studies. In the first level of screening, titles and abstracts were reviewed and potentially relevant articles were retrieved and assessed for inclusion. The final selection of full-text articles was based on the inclusion criteria presented in Table 1. This report is meant to provide a summary of the safety of reprocessing and does not provide a comprehensive list of device reprocessors in Canada or recommend any specific methods of reprocessing medical devices.

Table 1: Selection Criteria

Criteria

Description

Population

Patients of any age in contact with semicritical or critical medical device

Intervention

Reprocessed single-use semicritical or critical medical device

Comparator

New single-use semicritical or critical medical device

Outcomes

Patient safety (e.g., adverse events, infections, mortality, hospitalizations)

Study designs

Health technology assessments, systematic reviews, randomized controlled trials, nonrandomized studies

Exclusion Criteria

Articles were excluded if they did not meet the selection criteria outlined in Table 1. Articles were also excluded if they evaluated a single-use device that was reused and not reprocessed between uses according to a described reprocessing protocol, whether in-hospital or by a third-party. Articles were excluded if they did not report patient-related outcomes. Systematic reviews were excluded if they did not provide an analysis of patient-related outcomes.

Critical Appraisal of Individual Studies

The included publications were critically appraised by 1 reviewer using the Downs and Black checklist13 for randomized and nonrandomized studies as a guide. Summary scores were not calculated for the included studies; rather, the strengths and limitations of each included publication were described narratively.

Summary of Evidence

Quantity of Research Available

A total of 770 citations were identified in the literature search. Following screening of titles and abstracts, 738 citations were excluded, and 32 potentially relevant reports from the electronic search were retrieved for full-text review. One potentially relevant publication was retrieved from the grey literature search for full-text review. Of these potentially relevant articles, 25 publications were excluded for various reasons, and 8 publications met the inclusion criteria and were included in this report. These comprised 1 randomized controlled trial (RCT), and 7 nonrandomized studies. Appendix 1 presents the PRISMA14 flow chart of the study selection. Additional references of potential interest are provided in Appendix 5.

Summary of Study Characteristics

The literature search identified 8 publications meeting the inclusion criteria.2,15-21 Two publications describing a systematic literature review process conducted in Canada in 1994 were excluded because they focused on economic outcomes and did not provide any synthesis or analysis of patient-related outcomes.1,12

Additional details regarding the characteristics of included publications are provided in Appendix 2.

Study Design

Two prospective cohorts with historical controls were published in 199715 and in 2019,2 and 1 RCT was published in 2004.16 The rest of the included studies were prospective cohort studies published in 1994,17 1998,18 2018,22 2021,20 and 2022.21

Country of Origin

The included studies were conducted in the US,15 Canada,17Brazil,20 Iran,21 Turkey,16,18Portugal,22 and the UK.2

Patient Population

Browne et al.15 enrolled 107 patients scheduled for percutaneous transluminal coronary angioplasty (PTCA) at a single US institution. The mean age of the patients was 64 years, and 56% were male. The indication for coronary angioplasty was stable coronary insufficiency in 69 patients, unstable angina in 22 patients and acute myocardial infarction (MI) in 16 patients.

Plante et al.17 evaluated 693 patients with 853 lesions who underwent coronary angioplasty at 1 of 2 study centres. One study centre used new single-use catheters (n = 373 patients with 452 lesions), while the other used reprocessed single-use catheters (n = 320 patients with 401 lesions). The baseline patient and angiographic characteristics of the 2 groups were comparable except for a higher incidence of unstable angina (70% versus 57%. P < 0035) and nitroglycerin infusions (25% versus I5%) in patients at the reuse centre. The mean age in the single-use group was 60 years, and 73% were male. The mean age in the reprocessed group was 61 years, and 72% were male.

Zacharias et al.20 included 590 endolaser probes used during vitrectomy surgeries performed in a single institution, of which 375 (63.56%) were original and 215 (36.44%) were reprocessed. No details about the patients were provided.

The study population in Jokar et al21 included all candidates for elective laparoscopic cholecystectomy from August 2018 to August 2020. Detailed inclusion and exclusion criteria were provided. The reason for surgery was acute or chronic cholecystitis, biliary colic, or a group described as polyps, stones, or wall thickening. They evaluated 473 patients receiving a new port (n = 215) or a reprocessed disposable port (n = 258) in laparoscopic surgery. The mean age of patients in the new port group was 45.89 versus 46.29 years in the reprocessed port group. In the new port group, 68.8% of the patients were female, compared to 72.1% of the patients in reprocessed port group.

Marins et al.22 included all patients undergoing a wide variety of surgical interventions in a single year (2014) at their institution using specific surgical scissors and a suture machine. The mean age in the reprocessed device group (n = 316) was 56.7 years and 56.6 years in the nonreprocessed group (n = 417). There were 184 (58.2% females in the reprocessing group and 251 [60.2%]) in the nonreprocessed group.

Gundogdu et al.18 evaluated 45 patients (30 female, 15 male) who underwent laparoscopic cholecystectomy due to cholecystitis. The mean age of the patients was 48 years (range 21 to 71 years). Thirty patients were assigned to the reprocessed plastic trocar group, and 15 were assigned to the new plastic trocar group.

In the RCT by Colak et al.,16 of 125 consecutive patients with symptomatic cholelithiasis were randomly assigned to undergo laparoscopic cholecystectomy with either new (group 1) or reprocessed (group 2) disposable laparoscopic instruments. Group 1 consisted of 15 men and 47 women whose mean age was 51 years, and group 2 comprised 12 men and 51 women with a mean age of 52.

Leung et al.2 evaluated 100 consecutive patients undergoing elective atrial fibrillation (AF) ablation with a remanufactured circular mapping catheter. All cases were propensity matched to cases performed using a previously unused circular mapping catheter of the same model selected from a database of 806 patients receiving AF ablations over the previous 4 years. There were an equal number of males in each group (n = 68), and the average age in the remanufactured group was 67.1 years versus 65.8 years in the new catheter group.

Interventions and Comparators

For coronary angioplasty procedures, the Browne et al.15 study reused single-use PTCA balloon catheters shipped to a central facility and decontaminated, cleaned and tested for endotoxin using the Limulus amebocyte lysate gel clot method. Physical testing and quality assurance were performed. Standard angioplasty was then performed using these reprocessed devices. The comparison group was a case-matched control group of 108 patients evaluated retrospectively from an institutional database of patients undergoing PTCA with new catheters. In the Plante et al.17 study, 1 study site used new balloon catheters, and 1 study site employed previously used single-use balloon catheters subject to a strict reprocessing protocol described in detail. Standard angioplasty was performed using these devices.

Zacharias et al.20 included 590 endolaser probes used during vitrectomy surgeries of which 375 (63.56%) were new and 215 (36.44%) were reprocessed following a formal in-hospital reprocessing procedure that was detailed in the publication.

Jokar et al.21 evaluated new single-use surgical ports versus disposable ones previously used for other patients, cleaned and resterilized based on the hospital protocol. A detailed description of the reprocessing protocol was not provided.

Marins et al.22 evaluated a specific brand of single-use surgical scissors and suture machine, either new, or reprocessed, with the reprocessing carried out by a certified reprocessor. The devices can be reprocessed up to 2 times.

Colak et al.16 evaluated new disposable laparoscopic instruments versus reprocessed disposable laparoscopic instruments reused after high-level disinfection by alkalinized 2% glutaraldehyde solution following a detailed in-hospital reprocessing protocol. Gundogdu et al.18 evaluated new single-use plastic trocars versus reprocessed single-use plastic trocars for laparoscopic cholecystectomy, reprocessed according to a detailed in-hospital reprocessing protocol.

Leung et al.2 evaluated using third-party remanufactured single-use circular mapping catheters compared to new catheters for AF ablation. The remanufacturing process addressed both structural and functional integrity and a disinfection protocol.

Outcomes

Browne et al.15 evaluated patient temperature and white blood cell count (WBC) before and 24 hours after the PTCA procedure to screen for pyrogenic reactions. All patients were followed until hospital discharge for evidence of subsequent MI or requirement for emergent percutaneous or surgical revascularization of the target vessel. Long-term follow-up was not performed. The authors also reported mean procedure time, mean fluoroscopy time, mean dye volume, and the average number of balloons used per lesion, though it is unclear what impact these might have on patient outcomes. The need for additional procedures, as well the incidence of MI and mortality, along with the angiographic failure rate was also reported in the reprocessing group. Still, it was not reported for the comparison group.

Plante et al.17 reported procedure duration, fluoroscopy time and volume of contrast medium but did not report how these might impact patient outcomes. Temperatures were recorded 1 hour before angioplasty and every 8 hours during the first 24 hours after the procedure (starting 1 hour after angioplasty) and at least once daily until hospital discharge. Blood was drawn to measure creatine kinase levels before and at 8 and 24 hours after the procedure. Angiographic success was defined as lesional residual stenosis of less than 50%, as determined by visual assessment. Clinical success was defined as an angiographically successful angioplasty of all attempted lesions without any procedure-related in-hospital adverse clinical event, defined as death, MI, stroke, emergency angioplasty or bypass surgery. Clinical failure was designated when all attempted lesions could not be dilated successfully.

Zacharias et al.20 evaluated the functionality of endolaser probes in vitrectomy surgeries, with malfunction defined as the nonvisualization of the burns during intraoperative use that required the use of another probe. The authors did not describe malfunction as being related to patient outcomes. They also reported the frequency of infectious endophthalmitis. The authors describe infectious endophthalmitis as a rare but serious complication related to unfavourable visual outcomes.

Jokar et al.21 reported the incidence of port-site infection (PSI), both superficial and deep infection, through patient self-assessment obtained by telephone call using a researcher-made checklist of infection assessment standards at intervals of 3 to 5 days, and 5 to 14 days, and through physical examination in the hospital on day 14. Superficial infection was described as serous wound drainage, pain and tenderness, redness and warmth, edema and fever, while deep infection was described as suppuration, spontaneous dehiscence, and abscess.

Marins et al.22 reported on the duration of surgical intervention, postoperative infection incidence, antibiotic consumption, reoperations, length of hospital stay, in-hospital mortality, and rehospitalization rate (complications 30 days after surgical intervention).

Gundogdu et al.18 reported the incidence of infection at the wound site and intra-abdominally.

Colak et al.16 evaluated operating time, postoperative pain, duration of intramuscular analgesic administration, hospital stay, complications, and patient satisfaction. Operating time was defined as the time from the first incision to the placement of the last suture, while postoperative pain was assessed 6-hours after the operation according to a linear analogue scale (LAS) that ranged from 0 (no pain) to 10 (worst pain imaginable). The patients were also asked to indicate their overall level of satisfaction (1 = poor, 4 = excellent) at 1 week after cholecystectomy.

Leung et al.2 primarily evaluated device-related outcomes which could potentially affect patient outcomes, including catheter-related complications during or after the procedure, ease of handling the catheter, failure of electrodes to record electrograms or to stimulate appropriately, failure of communication with the electro-anatomic mapping system, and physical defect or deformation of the catheter on inspection after use. Indirect markers of catheter function were also measured, including procedure duration and fluoroscopy duration, as well as complications, including any major adverse cardiovascular/cerebrovascular events (MACCE), vascular injury, or cardiac tamponade.

Summary of Critical Appraisal

Prospective Cohort Studies with Historical Controls

The Browne et al.15 prospective cohort study with historical controls demonstrated serious reporting issues, external and internal validity issues. Very limited outcomes were reported for the historical comparison group. The included participants were selected from the same hospital, but it is not described whether they are consecutive, or whether any potential participants might have been excluded, limiting the generalizability and validity of the study results. The retroactive comparison group was not well-described, and a list of potential confounders was not provided. Most outcomes were not reported with statistical significance or statistical ranges, including patient safety outcomes. Follow-up for fever, chills and elevated WBC was reported only up to 24 hours postprocedure, and therein patients were followed only until hospital discharge. There was no long-term follow-up. The study was powered to detect a 5% difference in the angiographic failure rates of new and reused balloons, yet the failure rate for the comparison group did not appear to be provided.

Leung et al.2 conducted a well-reported prospective cohort study with historical controls. Loss to follow-up was not reported, a power calculation was not described, and most importantly, there are bias and confounding concerns due to the lack of randomization and blinding, which is typical of observational studies. Also, as the study used historical controls, the treatment groups were not recruited over the same period, which means there could be differences between patients, providers, or other factors that may have impacted the observed outcomes.

Prospective Cohort Studies

The Plante et al.17 study was a prospective cohort study evaluating balloon catheters for coronary angioplasty at 2 sites in the same province in Canada, with 1 site using reprocessed single-use catheters and 1 site using new catheters. While the 2 sites were described as performing a comparative number of angioplasty procedures, having similar case selection, and comparable operator experience and skills, the use of a control group at a second centre has the potential for differences in patient groups, practice patterns and other factors that could confound the observed results. At baseline, there were statistically significant differences in the incidence of unstable angina (and related IV heparin and nitroglycerin use before the procedure) between the 2 centres, which may suggest that more high-risk lesions were treated in the reprocessed device group. Therefore, the observed outcomes (such as longer procedure duration, adverse events, etc.) may have been at least partly due to differences in the groups at baseline. Overall, the Plante et al.17 study was well-reported, but demonstrates issues with external validity and internal validity, particularly confounding.

Zacharias et al.20 reported on a prospective cohort study comparing reprocessed single-use endolaser probes with new probes in vitrectomy surgeries. The study demonstrated issues with reporting, external validity, and internal validity. There was no description of the patient groups or any indication of how the patients were selected or assigned to treatment. Because of the flaws in the study design there is considerable likelihood of bias, confounding, and lack of generalizability, limiting confidence in the reported results.

Jokar et al.21 conducted a well-reported prospective cohort study evaluating the incidence of PSIs in patients undergoing laparoscopic cholecystectomy. There was no description of blinding of patients or outcome assessors or allocation concealment. They did report quasi-randomized allocation to treatment groups (if the last digit of the ID patient was odd, the patient was selected for the disposable port group, and if it was even, the patient was selected for the reprocessed port group), which may have limited bias to some extent. There appears to be a small loss to follow-up of 3 patients that was not reported or addressed by the authors, though this small number is unlikely to have impacted the observed outcomes. Though adverse events were reported, they were largely determined through patient self-report (which may not be as objective as if assessed by a health care professional using a validated outcome measure). Follow-up may not have been sufficiently long-term to assess all adverse events.

Marins et al.22 conducted a well-reported prospective cohort study evaluating surgical scissors and suture machines. Although the study scored well in terms of external validity and therefore generalizability, and also in many aspects related to confounding and bias, it had a few flaws typical of observational studies, including lack of randomization of patients to each intervention arm with assignment concealment, and there was no blinding of patients and outcome assessors described. Length of follow-up and loss to follow-up were not reported, and a power calculation was not performed. Without a power calculation, it is unclear whether nonsignificant differences between groups were real or related to the underpowering of the study to detect a difference between groups.

Gundogdu et al.18 conducted a prospective cohort comparing new versus reprocessed single-use plastic trocars. The study demonstrated serious issues with reporting and external validity, and internal validity, with regards to bias and confounding. There was very little focus on patient outcomes. The primary outcome of interest seemed to be on culturing bacteria on surgical equipment and at the patient's surgical site rather than on actual patient outcomes such as infection.

Randomized Controlled Trials

Colak et al. (2004)16 conducted an RCT evaluating the reuse of disposable laparoscopic instruments compared with new use. The RCT was well-reported, and scored very well with regard to external validity, and internal validity, both bias and confounding, meaning that the study results are generalizable to similar populations, and that 1 can have confidence in the results of the study. The only issues noted were that the proportion of patients that agreed to participate in the study was not stated, and it is not known if blinding of outcome assessors occurred, though patients and surgeons were blinded. No power calculation was described, which may have affected the lack of observed differences between groups.

Additional details regarding the strengths and limitations of included publications are provided in Appendix 3.

Summary of Findings

Balloon Catheters

Infection

Browne et al.15 concluded that there was “no significant difference” in frequency of fever or WBC count (a measure of infection) between the reprocessing group and the new catheters group. No patient in the control group had chills, while 1 person in the reprocessing group had chills. Other causes explained the elevated WBC in the reprocessing group in 9 patients (MI in 4, hematoma in 3, steroids in 1, and urinary tract infection in 1).

Plante et al.17 reported that fever was noted in 3 patients in the reprocessing group (in association with urinary tract infection, pneumonia and the postoperative period after urgent bypass surgery) and in 1 patient in the single-use centre (flulike illness), and in all cases, fever did not appear to be related to the catheterization procedure.

Other Outcomes

Browne et al.15 reported that the mean procedure time was 67 minutes in the reprocessing group compared with 83 minutes in the comparison group. The mean fluoroscopy time for the procedures was 13 minutes in the reprocessing group compared with 18 minutes in the comparison group. The mean dye volume was 275 mL compared with 307 mL in the comparison group. The incidence of device failure was 7%, which the authors claim compares with a 10% rate reported in the literature for new balloon catheters. Device failure was not reported for the comparison group in the retrospective review. P values were not reported for any outcomes.

Plante et al.17 reported that the angiographic success rate in both the reprocessed and single-use groups was identical at 88%. The reprocessing site utilized more balloon catheters per lesion (mean 2.4 catheters in the reprocessing group versus 1.2 in the single-use group, P < 0.00001), and had a higher incidence of initial balloon failure (10.2% versus 3.3%, P < 0.0001). The reprocessing group exhibited longer procedure times (81 minutes versus 68 minutes, P < 0.0001) and utilized an increased volume of contrast medium. (201 mL versus 165 mL, P < 0.0001), and increased adverse events, particularly in patients with unstable angina (7.8% versus 3.8%, P < 0.25). Hospital stay was significantly greater in the reprocessing group (5.1 days versus 3.4 days, P < 0.0001). No statistically significant differences in clinical success and failure were reported between the 2 groups. Overall clinical success rate and the rate of clinical failure without adverse clinical events were comparable in the 2 groups. Still, the rate of clinical failure with adverse clinical events was significantly higher in the reuse group (7.8% versus 3.8%. P < 0.025).

Endolaser Probes

Zacharias et al.20 conducted a two-phase study, with the first phase reporting in vitro results for the feasibility, sterility and physical integrity of single-use Endolaser probe reprocessing in vitreoretinal surgery. The results of the first phase are not in scope for this report based on the inclusion criteria in Table 1 and therefore are not reported here. The second phase included 590 endolaser probes, of which 375 were original and 215 were reprocessed, and reported statistically significant functionality rates between groups. Among the original probes, 373 (99.47%) were functioning and 2 (0.53%) were nonfunctioning. Among the reprocessed ones, 201 (93.5%) were functioning and 14 (6.5%) were nonfunctioning (P < 0.001) for an odds ratio OR = 12.99 (P < 0.001). It is not clear if the functionality rates would impact patient outcomes. The frequency of infectious endophthalmitis was zero in both groups.

Surgical Ports

Jokar et al.21 reported that the incidence of PSI was significantly higher in the reprocessed port group than in the new port group at all evaluated time points up to 2 weeks after surgery. All PSIs were considered superficial, and no symptoms of deep infection were reported. Individuals in the disposable port group had a maximum of 2 symptoms of superficial infection, but there were usually more than 2 symptoms in the reprocessed port group.

Surgical Scissors and Suture Machines

Marins et al.22 reported that the differences in postoperative infection rate between the reprocessed and nonreprocessed group were not statistically significant. There was a wide range of differences in postoperative infection rate depending on the severity of the condition and the type of surgery. There were also no statistically significant differences in antibiotic consumption.

The duration of the surgical intervention was on average 2 hour and 23 minute (P = 0.161), and length of hospital stay was approximately 10 days in both groups (P = 0.881). Rehospitalization was required for 28 patients in the reprocessing group and for 33 patients in the nonreprocessed group (P = 0.678). There was no statistically significant difference in mortality (P = 0.396). Hospital mortality in the reprocessing group was 1.90%, resulting in 6 deaths, and it was 2.88% (12 deaths) in the nonreprocessing group. There were no reoperations in either group.

Laparoscopic Plastic Trocars

Gundogdu et al.18 focused mainly on nonpatient outcomes in a prospective cohort study. Nonpatient related outcomes are not reported here because they are out of scope. The only patient outcome reported was infection which was evaluated using Cruise’s criteria. None of the patients had infection at the wound site or intra-abdominally.

Laparoscopic Surgical Instruments

Colak et al.16 reported that there was no statistically significant difference between the new and the reprocessed disposable laparoscopic instrument groups in mean operating time, LAS, duration and amount of analgesic administration, or hospital stay. The total incidence of complications (3.2% vs 4.8%, P = 0.50) and infection rates (1.6% vs 3.2%, P = 0.57) were also similar. The median satisfaction rate of the patients was excellent in both groups (P = 0.87)

Circular Mapping Catheters

Leung et al.2 reported that the procedure duration was similar in the remanufactured group (178.9 minutes) and the matched new catheter control group (189.5), P = 0.16). There was a statistically significant higher overall fluoroscopy duration in the new catheter group (21.5 minutes) versus the remanufactured group (11.8 minutes), P < 0.0001). There were no minor or major complications in either group. There was 1 mapping catheter failure in the manufactured group versus no failures in the new use group.

Appendix 4 presents the main study findings.

Limitations

There is just 1 study conducted in Canada,17, and the results of the studies conducted in the other countries may not be generalizable to Canadian populations and settings. In addition, the study conducted in Canada was published in 1994, and it is unknown if the same surgical methods, device specifications, patient care protocols, and reprocessing standards would be relevant today. In fact, of the 8 included studies,415-18 were published before the year 2005, which potentially limits the applicability of at least half of the studies reported. As an example, in 1 study using historical controls, the authors suggested that the decreased fluoroscopy time observed in the reprocessed device group during angiography was likely due to progressive decline in reliance on fluoroscopy seen in their institution over time, rather than to the intervention.2

The body of evidence was appraised to be of very low to moderate quality, and there is just 1 high-quality RCT16 that evaluated the use of reprocessed SUMDs. However, the RCT was published in 2004, was conducted in Turkey, and evaluates SUMDs that were reprocessed using an in-hospital protocol, which could limit its generalizability to the Canadian population and Canadian health care settings, as well as setting where third-party reprocessing is utilized.

While 8 studies met the inclusion criteria in this Rapid Review, they mostly looked at different devices for different interventions, which limits the amount of evidence available to evaluate the patient outcomes for a specific device/intervention. Two studies evaluated balloon catheters for coronary angioplasty,15,17 1 evaluated a mapping catheter for AF,2 1 evaluated plastic trocars for laparoscopic cholecystectomy,18 1 evaluated disposable laparoscopic instruments for laparoscopic cholecystectomy,16 1 evaluated surgical ports in laparoscopic cholecystectomy,21 1 evaluated surgical scissors and suture machine across a wide variety of surgical interventions,22 1 study evaluated endolaser probe reprocessing in vitreoretinal surgery.20 Due to the heterogeneity across studies, including reported outcomes, it is difficult to make comparisons or draw conclusions about the safety of reusing SUMDs. The majority of studies did not report statistically significant impacts on patient outcomes.

None of the included studies evaluated semicritical use devices, and all of the included studies took place in surgical settings.

Finally, the majority of the studies employed an in-house reprocessing protocol, and not the presumably more rigorous third-party reprocessing protocol, which has the potential to impact study results. Protocols varied by institution, by device, and it is assumed, also by third-party processor.

Conclusions and Implications for Decision- or Policy-Making

Conclusions Based on the Available Evidence

This Rapid Review identified 8 studies evaluating the use of reprocessed SUMDs compared with new SUMDs, most of which employed a different reprocessing protocol, and evaluated different reprocessed SUMDs in patients undergoing different surgical interventions. Three studies2,15,19 used third-party reprocessing or remanufacturing of SUMDs, while the rest described in-hospital reprocessing protocols that varied by institution. Of the 8 included studies, only 2 studies evaluated the same device in the same population (balloon catheters for coronary angiography15,17) while the rest of the included studies evaluated a variety of devices across different populations. It is difficult to compare outcomes or draw conclusions on the safety of reprocessed SUMDs based on the existing body of evidence since the population under study, the intervention, the outcomes reported, and the method of reprocessing varies. In addition, most of the published evidence was of very low to moderate quality, and published in countries outside of Canada, with at least half of the evidence published before 2005, which raises concerns regarding generalizability and applicability. It is likely that device specifications, reprocessing protocols, and surgical approaches and patient care protocols have changed since some of these studies were published.

The 2 studies that reported on the same intervention (balloon catheters) in the same population (coronary angiography) came to different conclusions, with Plante et al17 reporting higher rates of adverse events and other procedure-related issues in the reprocessed SUMD group, while Browne et al15 described much more favourable results. However, Plante et al17 used an in-house reprocessing protocol, whereas Browne et al15 used a third-party reprocessor, and there may have been differences between the patients groups at baseline in each study. Three studies evaluated the same study population, patients undergoing laparoscopic cholecystectomy, but in each case evaluated a different intervention: disposable laparoscopic instruments,16 plastic trocars,18 and surgical ports.21 For laparoscopic cholecystectomy patients, the use of reprocessed surgical ports led to a significant increase in PSI,21 while the use of reprocessed plastic trocars and other disposable laparoscopic instruments reportedly did not increase infection rates.16,18 The majority of studies did not report statistically significant differences in patient outcomes with the use of reprocessed SUMDs. However, it is unclear whether the observed outcomes are due to the reuse of single-use devices, or to differences between the devices themselves, or reprocessing protocols.

Studies evaluating semicritical devices were not identified, and all included studies evaluated critical category devices in surgical settings. However, it may be reasonable to assume that there would not be greater risks to patient safety when considering use of reprocessed semicritical SUMDs than reprocessed critical SUMDs. Most of the included studies had limited follow-up, so long-term outcomes for many patients were not available. In addition, outcomes were reported in which the impact on patient outcomes was not made clear. For example, reported outcomes such as procedure duration, fluoroscopy duration, and device failure would likely have some impact on patient outcomes, but it was not always clear what these impacts might be.

In the included studies, it was not always clear how often a SUMD was reprocessed, which is an important consideration in any reprocessing policy. For balloon catheters specifically, Browne et al.15 reported that it is unknown how many cycles of restoration a single balloon angioplasty catheter can tolerate, and that data are required to evaluate the degradation of these products with each subsequent cycle. Browne et al.15 also reported that using simple methods of cleaning and decontamination followed by radiation sterilization does not offer enough of a guarantee of safety to permit reuse. Plante et al.17 have emphasized that before a policy of balloon catheter reuse is initiated, it is imperative to establish clear and rigorous guidelines for cleaning and sterilization, and with adequate monitoring and quality control in place at all stages. Browne et al.15 emphasized the important differences between in vitro evaluation and in vivo evaluation and stated that methods of restoration must be highly controlled and reproducible. Marins et al22 emphasize that a certified reprocessing process requires rigorous inspection, with demanding criteria and guidelines specific to each device. They point out that these stringent criteria allow the detection of any type of anomaly, which is not the case in original medical devices whose quality control is done by sampling.

The reuse of SUMDs also has important potential economic and environmental benefits, which have been evaluated elsewhere.9-12

Standards and Resources to Inform Medical Device Reprocessing

In community health care settings (defined as any location outside of an acute care hospital where health care is provided), reprocessing of critical and semicritical medical equipment/devices, when not performed according to current standards, has been linked with health care-associated infections and outbreaks.23 Therefore, a 2019 position statement for community health care settings published by Infection Prevention and Control Canada, states that critical and semicritical medical equipment/devices labelled as single-use must not be reprocessed and reused unless a licensed reprocessor does the reprocessing.23

The Canadian Standards Association has published a 2023 standard regarding Canadian medical device reprocessing for all health care settings (CSA Z314:23), superseding an earlier version published in 2018, but excludes reprocessing of SUMDs.24

In 2018, the FDA (FDA) reviewed and officially reported on the remanufacturing industry over the previous 2 decades and found no evidence that using remanufactured devices increased risk to health.2 The Medicines and Health care Products Regulatory Agency also permits remanufacturing single-use devices within strict guidelines.25

Suggestions for Future Research

There is a need for high-quality studies evaluating patient outcomes using the following reprocessed SUMDs in Canadian health care settings and populations, using standardized reprocessing protocols, to inform the existing body of evidence. Standardized measures should be employed to evaluate the desired patient outcomes, and the length of follow-up is an important consideration. Defining the number of reprocessing cycles that a particular SUMD can tolerate is also important.

References

1.Hailey D, Jacobs PD, Ries NM, Polisena J. Reuse of single use medical devices in Canada: clinical and economic outcomes, legal and ethical issues, and current hospital practice. Int J Technol Assess Health Care. 2008;24(4):430-436. PubMed

2.Leung LW, Evranos B, Grimster A, et al. Remanufactured circular mapping catheters: safety, effectiveness and cost. J Interv Card Electrophysiol. 2019;56(2):205-211. PubMed

3.Reprocessing of Single-Use Medical Devices: A 2015 Update. Ottawa (ON): CADTH; 2015: https://www.cadth.ca/reprocessing-single-use-medical-devices-2015-update. Accessed 2023 Oct 12.

4.Spaulding's Classification of Medical Equipment/Devices and Required Level of Processing/Reprocessing. Toronto (ON): Public Health Ontario; 2017: https://www.publichealthontario.ca/-/media/documents/c/2017/cds-spaulding-table.pdf?la=en. Accessed 2023 Sep 25.

5.Reprocessed Single-Use Medical Devices. Ottawa (ON): CADTH; 2023: https://www.cadth.ca/sites/default/files/pdf/htis/2023/RB1671%20Single-Use%20Devices%20Final.pdf. Accessed 2023 Sep 25.

6.Update: Notice to Stakeholders - Health Canada's Regulatory Approach to Commercial Reprocessing of Medical Devices Originally Labelled for Single Use. Ottawa (ON): Health Canada; 2016: https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices/activities/announcements/update-notice-stakeholders-regulatory-approach-commercial-reprocessing-medical-devices-originally-labelled-single-use.html. Accessed 2023 Sep 25.

7.Reusable & Single-Use Medical Devices Standards: Standards for the reprocessing of reusable medical devices and for the use of single-use medical devices in all health care facilities and settings. Edmonton (AB): Alberta Health; 2019: https://open.alberta.ca/dataset/fd371ac2-b2be-49ac-93ef-43865a0bc0fb/resource/56c1cd3c-b617-4d91-947d-3e0e4a68cd09/download/health-reusable-single-use-medical-devices-standards.pdf. Accessed 2023 Sep 25.

8.Best Practices for Cleaning, Disinfection and Sterilization of Medical Equipment/Devices In All Health Care Settings, 3rd edition Toronto (ON): Public Health Ontario; 2013: https://www.publichealthontario.ca/-/media/Documents/B/2013/bp-cleaning-disinfection-sterilization-hcs.pdf?rev=7a36d8e526644eb794a9ef066d97a120&sc_lang=en. Accessed 2023 Sep 25.

9.Keil M, Viere T, Helms K, Rogowski W. The impact of switching from single-use to reusable healthcare products: a transparency checklist and systematic review of life-cycle assessments. Eur J Public Health. 2023;33(1):56-63. PubMed

10.Lam K, Gadi N, Acharya A, Winter Beatty J, Darzi A, Purkayastha S. Interventions for sustainable surgery: a systematic review. Int J Surg. 2023;109(5):1447-1458. PubMed

11.Bolten A, Kringos DS, Spijkerman IJB, Sperna Weiland NH. The carbon footprint of the operating room related to infection prevention measures: a scoping review. J Hosp Infect. 2022;128:64-73. PubMed

12.Jacobs P, Polisena J, Hailey D, Lafferty S. Economic analysis of reprocessing single-use medical devices: a systematic literature review. Infect Control Hosp Epidemiol. 2008;29(4):297-301. PubMed

13.Downs SH, Black N. The feasibility of creating a checklist for the assessment of the methodological quality both of randomised and non-randomised studies of health care interventions. J Epidemiol Community Health. 1998;52(6):377-384. PubMed

14.Liberati A, Altman DG, Tetzlaff J, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. J Clin Epidemiol. 2009;62(10):e1-e34. PubMed

15.Browne KF, Maldonado R, Telatnik M, Vlietstra RE, Brenner AS. Initial experience with reuse of coronary angioplasty catheters in the United States. J Am Coll Cardiol. 1997;30(7):1735-1740. PubMed

16.Colak T, Ersoz G, Akca T, Kanik A, Aydin S. Efficacy and safety of reuse of disposable laparoscopic instruments in laparoscopic cholecystectomy: a prospective randomized study. Surg Endosc. 2004;18(5):727-731. PubMed

17.Plante S, Strauss BH, Goulet G, Watson RK, Chisholm RJ. Reuse of balloon catheters for coronary angioplasty: a potential cost-saving strategy? J Am Coll Cardiol. 1994;24(6):1475-1481. PubMed

18.Gundogdu H, Ocal K, Caglikulekci M, Karabiber N, Bayramoglu E, Karahan M. High-level disinfection with 2% alkalinized glutaraldehyde solution for reuse of laparoscopic disposable plastic trocars. J Laparoendosc Adv Surg Tech A. 1998;8(1):47-52. PubMed

19.de Sousa Marins B, Queiroz e Melo J, Logarinho Monteiro JL, Rente G, Teixeira Bastos P. Reprocessing of single-use medical devices: Clinical and financial results. Port J Public Health. 2019;36(3):150-156. https:// karger.com/ pjp/ article/ 36/ 3/ 150/289169/ Reprocessing -of -Single -Use -Medical -Devices. Accessed 2023 Sep 19.

20.Zacharias LC, da Silva Conci L, Megnis BP, et al. Safety and cost-effectiveness of single-use endolaser probe reprocessing in vitreoretinal surgery. Int J Retina Vitreous. 2021;7(1):22. PubMed

21.Jokar M, Larti N, Zarei M, et al. Comparative Study of the Incidence of Port Site Infection in Disposable Ports and Reprocessed Disposable Ports in Laparoscopic Cholecystectomy. Surg Laparosc Endosc Percutan Tech. 2022;32(6):650-654. PubMed

22.de Sousa Marins B, Queiroz e Melo J, Logarinho Monteiro JL, Rente G, Teixeira Bastos P. Reprocessing of single-use medical devices: Clinical and financial results. Port J Public Health. 2019;36(3):150-156.

23.POSITION STATEMENT: Reprocessing of Critical and Semi Critical Devices in Community Healthcare Settings. Winnipeg (MB): Infection Prevention and Control Canada. 2019. https://ipac-canada.org/photos/custom/Members/pdf/Reprocessing%20of%20Critical%20and%20%20%20Semi-Critical%20Devices%20in%20the%20Community_Pos.._6Nov2019%20(1)%20-%20FINAL.pdf Accessed 2023 Sep 25. 2019.

24.CSA Z314:23: Canadian medical device reprocessing in all health care settings. Toronto (ON): Canadian Standards Association; 2023: https://www.csagroup.org/store/product/2704392/. Accessed 2023 Sep 26.

25.Single-use medical devices: UK guidance on re-manufacturing. London (UK): Medicines and Healthcare Products Regulatory Agency; 2016: https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/534784/Remanufacture_SUD_guidance.pdf. Accessed 2023 Sep 26.

Appendix 1: Selection of Included Studies

Figure 1: Selection of Included Studies

There were 770 citations identified and 738 excluded, with 32 electronic literature and 1 grey literature potentially relevant full-text reports retrieved for scrutiny. In total 8 reports are included in the review.

Appendix 2: Characteristics of Included Publications

Note that this appendix has not been copy-edited.

Table 2: Characteristics of Included Primary Clinical Studies

Study citation, country, funding source

Study design

Population characteristics

Intervention and comparator(s)

Clinical outcomes, length of follow-up

Browne et al. (1997)15

US

Funding source:

Supported by Watson Clinic Center for Research, Inc., Lakeland, Florida.

Prospective cohort with historical controls

All patients scheduled for coronary angioplasty at a single institution

PTCA (Reprocessed):

N = 107 Mean age, years (SD): 64 (12)

(range: 29 to 87 years)

Male: 56%

Stable coronary insufficiency: 69

Unstable angina: 22 Acute MI: 16.

Comparator (new):

N = 108

Case-matched control group from retrospective database

Intervention:

Reprocessed PTCA balloon catheters using strict reprocessing protocol

Comparator:

New single-use PTCA balloon catheters

Outcomes:

Temperature, WBC, chills within 24 hour

Follow-up: Within 24 hours and until hospital discharge

No long-term follow-up performed

Plante et al.(1994)17

Canada

Funding source:

Supported in part by a grant from the Centre Hospitalier Universitaire de Sherbrooke (the reprocessed device centre)

Prospective cohort (2-site)

N = 693 patients with 853 lesions undergoing coronary angioplasty

Balloon catheter (reprocessed) group:

n = 320 patients with 401 lesions

Mean age: 61 +/− 11 years

72% male

Balloon catheter (new):

n = 373 patients with 452 lesions

Mean age: 60 +/− 11 years

73% male

Significant between-group differences:

Unstable angina

Reprocessed: n = 224 (70%)

Single use: n = 213 (57%)

P < 0.005

Nitroglycerin infusions

Reprocessed: n = 81 (25%)

Single use: n = 54 (15%)

P < 0.0005

Intervention:

Reprocessed balloon catheters using strict reprocessing protocol

Comparator:

New single-use balloon catheter

Outcomes:

  • Temperature

  • Procedure duration

  • Fluoroscopy time

  • Volume of contrast medium

  • Angiographic success rate

  • Clinical success rate

  • Clinical failure rate

Follow-up:

Temperatures were recorded 1 hour before angioplasty and every 8 hours during the first 24 hours after the procedure, and at least once daily until hospital discharge.

Zacharias et al. (2021)20

Brazil

Funding source:

Self-funded

Prospective cohort (single site)

Patients undergoing vitreoretinal surgery

Population characteristics not described

N = 590 endolaser probes

n = 375 new endolaser probes;

n = 215 reprocessed using a formal reprocessing protocol

Outcomes:

Probe malfunction

Incidence of infectious endophthalmitis

Follow-up:

1 year

Jokar et al. (2022)21

Iran

Funding: NR

Prospective cohort

N = 473 candidates for elective LC meeting the inclusion criteria

Reason for surgery:

  • Acute or chronic cholecystitis

  • Biliary colic

  • Polyps, stones, or wall thickening

New port: n = 215

Reprocessed disposable port: n = 258

Mean age:

New port: 45.89 years

Reprocessed disposable port: 46.29 years

Gender

New port: 68.8% female

Reprocessed disposable port: 72.1%

Intervention:

Reprocessed single-use surgical port, limited description of reprocessing according to hospital protocol

Comparator:

New single-use surgical port

Outcomes and follow-up:

PSI 3 to 5 days after surgery, 5 to 14 days after surgery, and overall, 3 to 14 days after surgery

Marins et al. (2018)22

Portugal

Funding: NR

Prospective cohort

All patients undergoing a wide variety of surgical interventions in a single year at a single institution

N = 733

Mean age:

Reprocessed device group (n = 316):

56.7 (± 17.3) years

Nonreprocessed device group (n = 417):

56.6 (± 16.7) years

Intervention:

Reprocessed single-use linear suture machine GIA CovidienTM and the Harmonic ACE® scissors reprocessed by a certified reprocessor

Comparator:

New single-use linear suture machine GIA CovidienTM and the Harmonic ACE® scissors

Outcomes:

  • Duration of surgical intervention

  • Postoperative infection incidence

  • Antibiotic consumption

  • Reoperations

  • Length of hospital stay

  • In-hospital mortality

  • Rehospitalization rate (complications 30 days after surgical intervention).

Follow-up:

NR but 30-day rehospitalization rate reported so assumed follow-up of 30 days

Gundogdu et al. (1998)18

Turkey

Funding: NR

Prospective cohort

N = 45 patients undergoing LC for cholecystitis

30 female, 15 male

Mean age: 48 years (range approximately21 to 71 years)

Reprocessed trocars: n = 30

New single-use trocars: n = 15

Intervention:

Plastic trocars reprocessed according to a detailed protocol, including after high-level disinfection by alkalinized 2% glutaraldehyde

Comparator:

New single-use plastic trocars

Outcomes:

Infection at the wound site or intra-abdominally

Follow-up:

Until surgical discharge on the 1st or 2nd postoperative day

Colak et al. (2004)16

Turkey

Funding: NR

RCT

N = 125 consecutive patients with symptomatic cholelithiasis

New DLIs: n = 62

Gender: 15 men and 47 women

Mean age: 51 ± 12.57 years

Reprocessed DLIs (n = 63):

Gender: 12 men and 51 women (P = 0.48)

Mean age: 52 ± 12.29 years (P = 0.39)

LC using new DLIs

LC using reprocessed DLIs

Reprocessed using high-level disinfection by alkalinized 2% glutaraldehyde solution according to a detailed reprocessing protocol

Outcomes:

  • Operating time

  • Postoperative pain

  • Duration of intramuscular analgesic administration

  • Hospital stay

  • Complications

  • Satisfaction of patient and surgeon

Follow-up:

6 months

Leung et al. (2019)2

Funding: One author reported funding from Biosense Webster and from Medtronic (not clear if funding was for this study)

Prospective cohort with historical controls

N = 100 consecutive patients undergoing elective AF ablation using a remanufactured catheter

Propensity matched case-controls from a database of 806 receiving AF ablation over the previous 4 years

Males: n = 68 in both groups

Average age:

Remanufactured:

67.1 ± 8.5 years

New

65.8 ± 9.1 SD

Remanufactured circular mapping catheter (third-party remanufacturer) at the first reuse cycle

New use circular mapping catheter

Outcomes:

  • Catheter-related complications during or after the procedure;

  • Ease of handling the catheter;

  • Failure of electrodes to record electrograms or to stimulate appropriately;

  • Failure of communication with the electro-anatomic mapping system;

  • Physical defect or deformation of the catheter on inspection after use

  • Procedure duration

  • Fluoroscopy duration

  • Major or minor adverse events, such as cardiovascular/cerebrovascular events (MACCE), vascular injury, or cardiac tamponade

Follow-up:

3 months

AF = atrial fibrillation; DLIs = disposable laparoscopic instruments; LC = laparoscopic cholecystectomy; NR = not reported; PTCA = percutaneous transluminal coronary angioplasty; RCT: randomized controlled trial.

Note: For all studies, outcomes clearly not related to patient outcomes, such as economic or in vitro outcomes, were not included here as they are out of scope.

Appendix 3: Critical Appraisal of Included Publications

Note that this appendix has not been copy-edited.

Table 3: Strengths and Limitations of Clinical Studies Using the Downs and Black Checklist13

Strengths

Limitations

Browne et al. (1997)15

Reporting

  • The objective of the study is clearly described

  • The intervention and comparator are described

  • Estimates of the random variability in the data for the main outcomes are provided, and important adverse events have been reported

External validity

  • The subjects asked to participate in the study appeared to be representative of the entire population from which they were recruited

  • It appears that the staff, places, and facilities where the patients were treated were representative of the treatment the majority of patients receive

Internal validity (bias)

  • Follow-up was the same for all patients

  • The statistical tests seemed appropriate

  • Outcome measures appeared to be valid and reliable

Internal validity (confounding)

  • The patients in both groups were recruited from the same population

Power

  • The study was powered to detect balloon failure rates

Reporting

  • Only some of the outcomes are described in the introduction/methods

  • The characteristics of the patient group are not clearly described. While we know the source, no inclusion/exclusion criteria are reported.

  • There is no description of the characteristics of the retrospective control group, though it is described as being case-matched, and therefore the distribution of potential confounders is unknown

  • Outcomes are not presented clearly, and many outcomes are presented which were not detailed in the introduction or methods

  • P values were not reported

  • Loss to follow-up was not reported

External Validity

  • The proportion of study participants that agreed to participate is not stated, nor is this known for the comparator group

Internal Validity (Bias)

  • It is not known if patients and outcome assessors were blinded to treatment assignment

  • Unplanned outcomes and analyses were reported

Internal Validity (Confounding)

  • The patients in each group were recruited over different time periods

  • The patients were not randomized and therefore no randomization concealment

  • There does not appear to be any adjustment for confounding since potential confounders are not described

Plante et al. (1994)17

Reporting

  • The objectives of the study, the main outcomes, description of the interventions, and the characteristics of the included patients were well-described

  • The distributions of principal confounders in each group of subjects to be compared clearly described

  • The main findings are clearly described, and estimates of variably are provided

  • It appears that all-important adverse events have been reported

  • Actual P values were reported

External validity

  • The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited

  • It appears that the staff, places, and facilities where the patients were treated were representative of the treatment the majority of patients receive

Internal Validity (Bias)

  • No unplanned analyses occurred and follow-up appeared to be the same between both groups

  • Statistical analyses seemed to be appropriate

  • The main outcomes measures used seemed to be reliable

Internal validity (confounding)

  • Patients in the different treatment groups were recruited over the same period of time

Reporting

  • Loss to follow-up was not reported

External Validity

  • The proportion of those asked who agreed to participate was not provided

Internal Validity (Bias)

  • No attempt to blind patients or outcome assessors was described

Internal Validity (Confounding)

  • The patients in the different intervention groups were not recruited from the same population

  • Patients were not randomized to intervention groups and randomization concealment did not occur

  • There were confounders between treatment groups that may not have been considered in the analysis (i.e., higher rates of unstable angina in the reprocessed group at baseline)

Power

  • No power calculation was described

Other: Very limited length of follow-up may not be adequate to assess outcomes

Zacharias et al. (2021)20

Reporting

  • The study objectives were clearly described

  • The main outcomes were described in the introduction

  • The interventions of interest are clearly described

External validity

  • It appears that the staff, places, and facilities where the patients were treated were representative of the treatment the majority of patients receive

Internal validity (bias)

  • Length of follow-up was the same for both groups

  • No unplanned analysis was presented

  • The statistical tests seemed appropriate

Internal Validity (Confounding)

  • The patients in different intervention groups were recruited at the same hospital and were recruited over the same period of time

Reporting

  • The characteristics of the patients included in the study are not described, nor is the distribution of potential confounders

  • P values were not reported

  • Estimates of random variability are not reported

  • It is not clear whether all potential adverse events have been reported

  • The outcomes were clearly described but it was not clear if other complications should have been reported due to device malfunction

  • It is unknown whether patients have been lost to follow-up

External Validity

  • It is unknown whether the subjects asked to participate in the study were representative of the entire population from which they were recruited

  • It is not known if the patients that were prepared to participate representative of the entire population from which they were recruited

Internal Validity (Bias)

  • It is not known if patients and outcome assessors were blinded to treatment assignment

  • The main outcome measures related to patient outcomes were not sufficiently described

Internal Validity (confounding)

  • Patients were not randomized and allocation concealment did not occur

  • It is not stated whether any adjustment was made for confounding, and potential confounders at not described

  • It is unknown if loss to follow-up occurred and was addressed if so

Power

  • No power calculation was described

Jokar et al. (2022)21

Reporting

  • Study objectives were clearly described

  • Outcomes were clearly described

  • Characteristics of the patients included in the study were clearly described

  • The interventions of interest were clearly described

  • The distributions of principal confounders in each group of subjects to be compared were clearly described

  • The main findings of the study were clearly described

  • It is assumed that all-important adverse events that may be a consequence of the intervention were reported

  • Actual P values were reported

External Validity

  • The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited, given any patient requiring LC was eligible (excepting detailed inclusion and exclusion criteria)

  • The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive

Internal Validity (Bias)

  • No unplanned analyses occurred

  • Length of follow-up was the same in both groups

  • The statistical tests used to assess the main outcomes appeared to be appropriate

  • The outcome measures were clearly described

Internal validity – confounding

  • Both groups were recruited from the same population

  • Both groups were recruited over the same period of time

  • There was adequate adjustment for confounding factors between treatment groups at baseline

  • Only 3 patients appeared to be lost to follow-up

Power

  • A very simple power calculation was performed

Reporting

  • The study does not provide estimates of the random variability in the data for the outcome

  • Though adverse events were reported, they were largely determined through patient self-report which is not ideal, and follow-up may not have been sufficiently long-term to assess all adverse events

  • Potentially 3 patients were lost to follow-up but loss to follow-up was not reported or addressed

External Validity

  • The percentage of subjects asked to participate in the study was not reported, and therefore it is not known if those that participated are representative of the entire population from which they were recruited

Internal Validity (Bias)

  • No attempt was described to blind study subjects or outcome assessors to the intervention they received

Internal Validity (Confounding)

  • Quasi-randomization was performed to assign patients to each group

  • The randomized intervention assignment was not described as being concealed from patients and would not likely be concealed for health care staff

Marins et al. (2018)22

Reporting

  • Study objectives were clearly described

  • Outcomes were clearly described

  • Characteristics of the patients included in the study were clearly described

  • The interventions of interest were clearly described

  • The distributions of principal confounders in each group of subjects to be compared were clearly described

  • The main findings of the study were clearly described

  • All-important adverse events that may be a consequence of the intervention were reported

  • Actual P values were reported and estimates of random variability were reported

External Validity

  • The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited, given any patient requiring LC was eligible

  • It appears that patients were automatically enrolled in the study, so the proportion that agreed is not a factor

  • The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive

     Internal Validity – Bias

  • No unplanned analyses occurred

  • Follow-up appeared to be the same between groups

  • Statistical analysis appeared to be appropriate

  • Main outcome measures used were valid and appropriate

Internal validity – Confounding

  • Both groups were recruited from the same population

  • Both groups were recruited over the same period of time

  • There was adequate adjustment for confounding factors between treatment groups at baseline

Reporting

  • Loss to follow-up was not reported

Internal Validity – Bias

  • It is not known if blinding of patients and outcome assessors occurred

Internal validity – Confounding

  • There was no randomization or intervention assignment concealment

Power

  • No power calculation was described

Gundogdu et al. (1998)18

Reporting

  • The objective of the study is clearly described

  • The interventions of interest are clearly described

  • It is assumed that all adverse events have been reported

External Validity

  • The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive

Interval Validity - Bias

  • An attempt made to blind study subjects to the intervention they received

  • No unplanned analysis occurred

  • Statistical tests appeared to be adequate

  • The main outcome measures used appeared to be accurate

Internal Validity – Confounding

  • Both groups were recruited from the same population

  • Both groups were recruited over the same period of time

Reporting

  • The main outcomes to be measured were not clearly described in the Introduction or Methods section

  • The characteristics of the patients included in the study were not clearly described with inclusion and exclusion criteria

  • The distributions of principal confounders in each group of subjects are only partially described

  • The findings are not clearly described and estimates of the random variability in the data for the main outcomes are not provided

  • Loss to follow-up was not reported

  • Actual P values were not reported

External Validity

  • It is not clear whether the subjects asked to participate in the study were representative of the entire population from which they were recruited

  • The proportion of those asked who agreed to participate was not stated

Interval Validity - Bias

  • It is unknown whether the outcome assessor was blinded to the intervention

  • No adjustment appeared to be made for different lengths of follow-up

Internal Validity – Confounding

  • Randomization and allocation concealment was not reported

  • It is not stated whether any adjustment was made for confounding

  • Loss to follow-up was not reported

Power

  • No power calculation was reported.

Colak et al. (2004)16

Reporting

  • Study objectives were clearly described

  • Outcomes were clearly described

  • Characteristics of the patients included in the study were clearly described

  • The interventions of interest were clearly described

  • The distributions of principal confounders in each group of subjects to be compared were clearly described

  • The main findings of the study were clearly described

  • All important adverse events that may be a consequence of the intervention were reported

  • Actual P values were reported and estimates of random variability were reported

  • Loss to follow-up was reported

External Validity

  • The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited

  • The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive

Internal Validity – Bias

  • No unplanned analyses occurred

  • Follow-up was the same between groups

  • Statistical analysis was appropriate

  • Main outcome measures used were valid and appropriate

  • Blinding of patients to treatment allocation occurred

Internal Validity – Confounding

  • Patients were randomized to treatment groups and allocation concealment was implemented (only the nurse knew the allocation)

  • Both groups were recruited from the same population

  • Both groups were recruited over the same period of time

  • There was adequate adjustment for confounding factors between treatment groups at baseline

  • The proportion lost to follow-up was too small to affect the main findings

External Validity

  • The proportion of patients that agreed to participate was not stated

Internal Validity – Bias

  • It is not known if blinding of outcome assessors occurred

Power

  • No power calculation was described

Leung et al. (2019)2

Reporting

  • Study objectives were clearly described

  • Outcomes were clearly described

  • Characteristics of the patients included in the study were clearly described

  • The interventions of interest were clearly described

  • The distributions of principal confounders in each group of subjects to be compared were clearly described

  • The main findings of the study were clearly described

  • All important adverse events that may be a consequence of the intervention were reported

  • Actual P values were reported and estimates of random variability were reported

External Validity

  • The subjects asked to participate in the study seemed to be representative of the entire population from which they were recruited

  • The staff, places, and facilities where the patients were treated seemed to be representative of the treatment the majority of similar patients would receive

Internal Validity – Bias

  • No unplanned analyses occurred

  • Follow-up was the same between groups

  • Statistical analysis was appropriate

  • Main outcome measures used were valid and appropriate

Internal validity – Confounding

  • Both groups were recruited from the same population

  • There were no confounding factors between treatment groups at baseline

Reporting

  • Loss to follow-up was not reported

External Validity

  • The proportion of those asked who agreed to participate in the study was not stated.

Interval Validity - Bias

  • Blinding of patients to treatment allocation and did not occur

  • Blinding of outcome assessors was not described

Internal Validity – Confounding

  • Patients were not randomized to treatment groups and allocation concealment was not implemented

  • Both groups were not recruited over the same period of time

Power

  • No power calculation was described

Appendix 4: Main Study Findings

Note that this appendix has not been copy-edited.

Table 4: Summary of Findings

Study citation and study design

Type of reprocessed device

Reprocessed outcomes

New device outcomes

Notes

Browne et al. (1997)15

Retrospective case-control

PTCA Balloon Catheter

Fever: n = 11

Elevated WBC: n = 12

Chills: n = 1

Mean procedure time: 67 +/− 30 minutes

Mean fluoroscopy time: 13 +/− 10 minutes

Mean dye volume: 275 +/− 125 mL

Average balloons per lesion: 1.5 +/− 0.7

Angiographic failure rate:

7% (10 /108, 95% CI, 2% - 12%)

Fever: n = 12

Elevated WBC: n = 14

Chills: n = 0

Mean procedure time:

83 +/− 49 minutes

Mean fluoroscopy time:

18 +/− 15 minutes

Mean dye volume:

307 +/− 157 mL

Average balloons per lesion: 1.6 +/− 0.6

Angiographic failure rate:

NR

P values were not reported

Angiographic failure rate reportedly comparable to the 10% rate seen with new balloons in other studies but not reported for comparison group.

Plante et al. 199417

Balloon catheter

Angiographic success rate: 88%

Clinical failure with adverse clinical events:

n = 25 (7.8%)

Balloon catheter use per lesion:

Mean 2.4 +/− 1.5 catheters

Initial balloon failure: 10.2%

Procedure times:

81 +/− 41 minutes

Volume of contrast medium:

201 mL +/− 86

Adverse clinical events:

7.8%

Length of hospital stay:

5.1 +/− 5.7 days

Fever: n = 3

Angiographic success rate: 88%

Clinical failure with adverse clinical events:

N = 14 (3.8%)

P < 0.025

Balloon catheter use per lesion:

Mean 1.2 +/− 0.5 catheters

P < 0.00001)

Initial balloon failure:

3.3%; P < 0.0001

Procedure times:

68 +/− 32 minutes,

P < 0.0001

Volume of contrast medium:

165 +/− 61 mL

P < 0.0001)

Adverse clinical events:

3.8%, P < 0.025

Length of hospital stay:

3.4 +/− 2 to 8 days; (P < 0.0001)

Fever: n = 1

Fever in 3 patients in the reprocessing group was in association with urinary tract infection, pneumonia and in the postoperative period after urgent bypass surgery, not considered to be procedure related.

Fever in in 1 patient in the single-use centre (flulike illness) also not thought to be procedure related.

No statistically significant differences in clinical success and clinical failure without adverse events were reported between the 2 groups.

Zacharias et al. (2021)20

Endolaser Probe

Functionality of device

n = 201 (93.5%) functioning

n = 14 (6.5%) were nonfunctioning

Infectious endophthalmitis

n = 0

Functionality of Device

n = 373 (99.47%) functioning

n = 2 (0.53%) nonfunctioning

(P < 0.001)

OR = 12.99 (P < 0.001)

Infectious Endophthalmitis

n = 0

It is not clear whether functionality of the device is related to patient outcomes.

Jokar et al. 202221

Surgical port

PSI incidence at days 3 to 5:

6.6%

PSI incidence at days 5 to 14:

2.7

Overall PSI:

8.5%

PSI incidence at days 3 to 5:

1.9%

P = 0.013

PSI incidence at days 5 to 15:

0

P = 0.018

Overall PSI:

1.9%

P = 0.002

All PSIs in both groups were considered superficial, and no symptoms of deep infection were reported.

Individuals in the disposable port group had a maximum of 2 symptoms of superficial infection, but in the reprocessed port group, there were usually more than 2 symptoms.

Marins et al. (2018)22

Surgical scissors and suture machine

Postoperative infection rate

12.3%

Consumption of antibiotics

(defined daily dose)

4.10 ± 7.63

Length of Hospital Stay

9.55 ± 8.92 days

Duration of Surgery

2.33 ± 1.53 hours

Rehospitalization

n = 33 (7.91%)

Mortality

n = 6 (1.90%)

Reoperations

n = 0

Postoperative Infection Rate

13.4%

P = 0.664

Consumption of antibiotics

(defined daily dose)

5.61 ± 15.21

P = 0.319

Length of Hospital Stay

10.36 ± 12.00 days

P = 0.881

Duration of Surgery

2.45 ± 1.52 hours

P = 0.161

Rehospitalization

n = 28 (8.86%)

P = 0.678

Mortality

n = 12 (2.88%)

P = 0.396

Reoperations

n = 0

There was a wide range differences in postoperative infection rate depending on the severity of the condition and the type of surgery.

Mortality differed between patients in 2 surgery types, described as major procedures in the small and/or large intestine with complications (P = 0.044) and in major procedures in the stomach, esophagus, duodenum, small intestine, and/or large intestine, with complications (P = 0.141).

Gundogdu et al. (1998)18

Plastic trocar

Wound and intra-abdominal infection

N = 0

Wound and intra-abdominal infection

N = 0

Bacterial cultures on surgical equipment, disinfection solution, and various patient wound/bile locations are not reported here.

Colak et al. (2004)16

DLIs

Mean operating time (minutes):

43.88 ± 16.29

Mean LAS score

4.80 ± 1.07

Duration of Postop Analgesia (days)

0.90 ± 0.81

Amount of analgesic required (mg)

67.85 ± 61.30

Hospital stay (days)

1.30 ± 0.75

Conversion rate (%)

2 (3.2)

Infection

2 (3.2)

Other Complications

1 (1.6)

Mean operating Time (minutes)

42.66 ± 13.92, P = 0.65

Mean LAS score

4.95 ± 1.07, P = 0.54

Duration of Postop Analgesia (days)

0.88 ± 0.68, P = 0.89

Amount of analgesic required (mg)

65.32 ± 49.86, 0.49

Hospital stay (days)

1.53 ± 2.61, P = 0.50

Conversion rate (%)

1 (1.6)

Infection

1 (1.6%)

Other Complications

1 (1.6), P = 0.31

The satisfaction of the surgeon with the instruments was excellent in 46 patients (74.2%) and good in 16 patients (25.8%) of the new DLI group vs excellent in 41 patients.

(65.1%) and good in 22 patients (34.9%) of the reprocessed DLI group (P = 0.27).

The median satisfaction rate of the patients was excellent in both groups (P = 0.87).

Leung et al. (2019)2

Circular mapping catheter

Procedure duration:

178.9 ± 51.3 minutes

Fluoroscopy duration

11.8 ± 9.6 minutes, respectively,

Mapping catheter failure:

n = 1

Major or minor complications:

n = 0

Procedure duration:

189.5 ± 55.3 minutes, P = 0.16

Fluoroscopy duration:

21.5 ± 13.5 minutes, P < 0.0001

Mapping catheter failure:

n = 0

Major or minor complications:

n = 1 (cardiac tamponade)

Cardiac tamponade considered to be not procedure-related.

Decreased fluoroscopy time observed in the reprocessed device group during angiography was likely due to progressive decline in reliance on fluoroscopy seen in their institution.

DLIs = disposable laparoscopic instruments; LAS: linear analogue pain scale; NR = not reported; PSI = port-site infection; PTCA = percutaneous transluminal coronary angioplasty.

Note: For all studies, outcomes definitely not related to patient outcomes, such as economic and in vitro outcomes, were not included here as they are out of scope.

Appendix 5: References of Potential Interest

Reuse Without Processing

Saadat SH, Shepherd S, Van Asseldonk B, Elterman DS. Clean intermittent catheterization: Single use vs. reuse. Can Urol Assoc J. 2019;13(2):64-69. PubMed

Zabaleta-Del-Olmo E, Vlacho B, Jodar-Fernandez L, et al. Safety of the reuse of needles for subcutaneous insulin injection: A systematic review and meta-analysis. Int J Nurs Stud. 2016;60:121-132. PubMed

Environmental Impact of Single-Use Devices

Anonymous. Reducing the environmental impact of surgery on a global scale: systematic review and co-prioritization with healthcare workers in 132 countries. Br J Surg. 2023;110(7):804-817. PubMed